Site location:home >> product >> Machinerary for Pharmaceutical &Medical Devices Industry >>  >> 
Stent Durability Tester
Stent Coating Durability Tester
Balloon Hyraulic Burst/Leak Tester
Thermal Bonding Machine
Powder Medicine Filling Machine
Tablet Visual Inspection System
Capsule Visual Inspection System
Tablet Print Inspection System
Heart Valve Durability Tester
Balloon Bonding Machine
Reflow Equipment
Product features
Ability to test samples of varying sizes of the same design at the same precisely controlled test parameters.

Product details
Product standard
 

Coating Durability Tester

Dynatek’s patented* Coating Durability Tester utilizes a Hyper Drive linear motor to produce an adjustable bellows compression which allows for physiologically relevant stent loading conditions for up to 5 test samples from physiological to accelerated testing rates of up to 150Hz.

Dynatek’s laser particle detection system allows accurate real-time counting, sizing, and data logging of shed particles from each drug-eluting stent during testing, giving a time course of shed particulates from each sample.

Features and Capabilities

  • FDA accepted with appropriate documentation.
  • Ability to test up to 5 samples simultaneously under precisely controlled test parameters. **
  • Ability to test samples of varying sizes of the same design at the same precisely controlled test parameters.
  • Patented* system used to detect and identify, in real time, the number and size of shed particles from each sample.
  • Advanced universal controller with unmatched data collection ability.
  • Data logging to a TDMS file which can be accessed by Excel or other software such as DIAdem and Matlab, for managing, analyzing, and reporting the test data collected.
  • Universal controller allows for future expansion of the CDT-5 system to accommodate future host computer technology.
  • Future upgrades with advance SCADA technology allow
  • Unique redundant filtration system that allows for filter changes without test interruption.
  • Frequency, temperature, and pressure controls give the user the ability to establish, maintain, and recall required test parameters.
  • Ability to monitor sample tube pressures at all test frequencies.
  • Long life electromagnetic Hyper Drive linear motor.
  • Design allows for easy access to test samples.
  • Future upgrades with advance SCADA technology allow complete lab monitoring of all machines from a single host computer.

*US Patent # 7,621,192 B2
**Number of test samples dependent on diameter and compliance of sample tubes.

 

 

Product Specification
Description
Specifications
Additional Information
Number of test samples:
5 test samples
5 test samples standard
Tube configuration:
Straight, curved, or bifurcated
Additional special configurations upon request
Tube length (straight) :
adjustable to ≤ 215mm
Curved samples would
be longer
Tube diameter range:
1mm to ≤ 9mm
 
Typical distention:
0.3% to 18% of tube ID
Distention is dependent
on tube size
Full physiological
range possible
Maximum
distention:
Limited by tube displacement and construction
 
Testing fluid:
PBS or distilled water
Not limited to
distilled water
Fluid temperature:
 ambient to ≤ 45⁰C
Temperature control
accurate to
+/- 0.3° C of set temperature
Flow rate:
100 mL/min standard
10-210 mL/min
(with particle counter recalibration)
Test monitoring:
Stent deflection, tube deflection, pressure
Stent deflection requires high speed camera (best method)
Automatic test control method:
Tube distention
With optional laser micrometer

Particle Counter Specification
Description
Specifications
Additional Information
Number of flow paths:
5
5 Individual
isolated paths
Number of particle counters:
5
Individual particle counters
for each test sample
Particle size range:
7-1000 micron
May require different
calibration
Particle count range:
15,000 p/mL
@ 10 microns
≥8,000 p/mL above 150 microns
Particle bin range:
2-6 bins
Individually configurable
size ranges
Particle counter calibration standard:
USP <788>
 
Filter assembly:
0.2 micron
Included parallel filters
allow easy swap
without test interruption

Dimensions and Power Requirements
Module
Dimensions
Power Requirement
Weight
       
CDT-5 Bench Top system
48”W x 60”D
x 48”H
120 VAC 12 A
300Lbs.
(complete system)

Available Add-ons
  • Individual filters for each test sample.
  • Various diameters of graft tube adapters for different test sample sizes.
  • Compact integrated mounting stand or benchtop configurations available.
  • Closed loop laser micrometer feedback allows for automatic control and monitoring of long-term test parameters.
  • Valves to allow individual samples to be removed or replaced without interrupting the test.
  • Very large diameter test samples possible with custom hardware.
  • Alternative cap tanks and fluid heaters for higher test temperatures.
  • UPS (Uninterruptable Power Supply) system.
  • Adapters for catheter deployment of stents on the tester.
  • Custom particle counter calibration for atypical particle size ranges.
  • Custom collection filters available to aid in chemical analysis of particles.
  • Other custom modifications at customer request.

 

 